A Performance Comparison of Commercially Available Strontium Extraction Chromatography Columns
Document Type
Article
Publication Date
11-30-2015
Publication Title
Journal of Radioanalytical & Nuclear Chemistry
Volume
307
Issue
3
First page number:
1825
Last page number:
1831
Abstract
Extraction chromatographic separations following a nuclear event will require both speed and accuracy. Ideally these separations will be carried out using commercially available products. Because 90Sr is a likely material to be used in a radioactivity dispersal device, commercially available products for the separation of strontium were tested to evaluate differences in their elution profiles that may affect elution procedures. Gravity flow columns show better resolution, but vacuum flow columns allow rapid processing and are still adequate for purification if multiple aqueous conditions can be employed in the elution strategy.
Keywords
Extraction chromatography; Strontium; Elution comparison
Repository Citation
McLain, D. R.,
Mertz, C. J.,
Sudowe, R.
(2015).
A Performance Comparison of Commercially Available Strontium Extraction Chromatography Columns.
Journal of Radioanalytical & Nuclear Chemistry, 307(3),
1825-1831.