Maintaining Radiation Exposures As Low As Reasonably Achievable (ALARA) for Dental Personnel Operating Portable Hand-Held X-Ray Equipment
Document Type
Article
Publication Date
8-1-2012
Publication Title
Health Physics
Volume
103
Issue
2
First page number:
S179
Last page number:
S185
Abstract
Clinical experience indicates that newly available portable hand-held x-ray units provide advantages compared to traditional fixed properly installed and operated x-ray units in dental radiography. However, concern that hand-held x-ray units produce higher operator doses than fixed x-ray units has caused regulatory agencies to mandate requirements for use of hand-held units that go beyond those recommended by the manufacturer and can discourage the use of this technology. To assess the need for additional requirements, a hand-held x-ray unit and a pair of manikins were used to measure the dose to a simulated operator under two conditions: exposures made according to the manufacturer's recommendations and exposures made according to manufacturer's recommendation except for the removal of the x-ray unit's protective backscatter shield. Dose to the simulated operator was determined using an array of personal dosimeters and a pair of pressurized ion chambers. The results indicate that the dose to an operator of this equipment will be less than 0.6 mSy y(-1) if the device is used according to the manufacturer's recommendations. This suggests that doses to properly trained operators of well-designed, hand-held dental x-ray units will be below 1.0 mSv y(-1) (2% of the annual occupational dose limit) even if additional no additional operational requirements are established by regulatory agencies. This level of annual dose is similar to those reported as typical dental personnel using fixed x-ray units and appears to satisfy the ALARA principal for this class of occupational exposures. Health PhyS. 103(Supplement 2):S179-S185; 2012
Keywords
operational topics; as low as reasonably achievable (ALARA); safety standards; x-ray machines
Repository Citation
McGiff, T. J.,
Danforth, R. A.,
Herschaft, E. E.
(2012).
Maintaining Radiation Exposures As Low As Reasonably Achievable (ALARA) for Dental Personnel Operating Portable Hand-Held X-Ray Equipment.
Health Physics, 103(2),
S179-S185.
http://dx.doi.org/10.1097/HP.0b013e318259fa29